Specifications HT-110
Continuous UV oxidation and conductivity difference measurement
0 to 1000 ppb C
Measuring range : 0 to 100/500/1000 ppb C

0 to 100 ppb C
0 to 500 ppb C
0 to 1000 ppb C
Within ±2% of the full scale
Number of outputs : 1
DC 4 to 20mA / DC 0 to 1V (option) : Input/Output isolated type
Max. load resistance : 600 Ω(min. load resistance : 100kΩ [with option selected])
Analog output range : Freely selectable within the measurement range
Output contents : TOC concentration
Number of outputs : 2
Alarm contact output (RLY1)
Contact type : Relay contact, SPDT
Contact rating : AC 125V 3A, DC 24V 3A (resistance load)
Contact function : Upper/lower limit action (ON/OFF control)
Output contents : Selectable between TOC concentration, Temperature,Conductivity, Hold, and Error
(Operation is reversed when Hold or Error is selected)
Alarm contact output (RLY2)
Contact type : Relay contact, SPST
Contact rating : AC 125V 3A, DC 24V 3A (resistance load)
Contact function : Upper/lower limit action (ON/OFF control)
Output contents : TOC concentration
Number of input point : 1
Conductivity : < 10μS/cm or lower (0 to 100 ppb: < 2μS/cm or lower)
Temperature : 5 to 99 ℃
Pressure : 0.03 MPa to 0.50 MPa
Flow rate : 20 to 200 mL/min
Temperature : 5 to 40℃
Humidity : 85% or lower
LC with backlight
Display content (TOC concentration, conductivity, temperature, alarm)
Protection : IP43 ; Paint color 5PB8/1; Epoxy coating (indoor-use panel installation type)
AC 100 to 240 V ±10% 50/60 Hz 80 VA (max.)
13 kg
CE marking, FCC Part 15